158. In fluoroscopy, when the operator of an x-ray machine is stationed directly behind the detector, the primary x-ray beam must be located within the surface area of the fluoroscopic screen while leaving an unexposed margin of at least 6 mm.
R.R.Q., 1981, c. P-35, r. 1, s. 158.